DxH 900 Hematology Analyser

DxH 900 Hematology Analyser

Product: DxH 900 Hematology Analyser

Features:

The DxH 900 hematology analyzer delivers accurate and high-confidence results on the first run. With its best-in-class hematology technology, this automated analyzer minimizes slide reviews, maximizes laboratory staff time, and provides significant cost savings. The DxH 900 offers scalable, customized solutions to address the hematology testing needs of high-volume labs.

Product Information:

 
  • Methodology: Enhanced Coulter Principle, VCS 360 with DataFusion
  • Throughput: Up to 100 samples per hour
  • Menu/Test Parameters: CBC: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV

    Differential: NE, LY, MO, EO, BA, NRBC, NE#, LY#, MO#, EO#, BA#, NRBC #, MDW (when enabled) Retic: RET, RET#, MRV, IRF  Body Fluids (Cerebrospinal, Serous, or Synovial): RBC, TNC

  • Sample Volume: 165 μL aspiration volume; 0.5 ml dead volume. Predilute: 50 μL whole blood into 200 μL of diluent
 
  • Power Requirements: 100 – 240 VAC,  50/60 Hz, Single phase with ground
  • Power Consumption: DxH 900 SPM: 520 W (1775 BTU/hour), Monitor: 35 W (120 BTU/hour), DxH Power Computer: 160 W (546 BTU/hour)
  • Noise Level: Less than or equal to 60 dBa
  • Atmospheric Pressure: 700 – 1,060 mbar
  • Operational Ambient Temperature: 18 – 32°C (64.4 – 89.6°F)
  • Humidity: 80% relative humidity (non-condensing) at 32° C (89.6° F)
  • Data Management:

    Up to 50,000 patient results with histograms, scatter plots, and demographics

    100,000 patient results with graphics for power computer

  • Weight and Dimensions: Depth: 83.82cm (33 in.) , Width: 72,6 cm (30 in.) Height: , 176.53 cm (69.5 in.) ,Weight: 292.11 kg (644lbs lbs) 
  • Quality Assurance: QC with Levy-Jennings graph; XB/XM for moving averages, daily check, intelligent quality monitoring, MRV, IRF, customizable calibration and QC reminders and alerts, auto export of QC
  • Radio Frequency & Interference Immunity: Meet the product level standards EN 61326-1 (Electrical Equipment for Measurement, Control, and Laboratory Use—EMC Requirements), and EN 61326-2-6 (Part 2, Specific Requirements for In-Vitro Diagnostic Devices)